An investigation on methods for free-field comparison calibration of measurement microphones

Publikation: Bidrag til bog/antologi/rapportKonferencebidrag i proceedingsForskningfagfællebedømt

Free-field comparison calibration of measurement microphones requires that a calibrated reference microphone and a test microphone are exposed to the same sound pressure in a free field. The output voltages of the microphones can be measured either sequentially or simultaneously. The sequential method requires the sound field to have good temporal stability. The simultaneous method requires instead that the sound pressure is the same in the positions where the microphones are placed. In this paper the results of the application of the two methods are compared. A third combined method is suggested.